Copyright © Shenzhen Fuyida Instrument Co., Ltd. All Rights Reserved. Site Map
Model | TM80-6 | TM100-4 | TM165-2 | |
Luck mode | Forced air supply circulation, horizontal air transport | |||
Performance | Temperature range | Room temperature +30~200°C | ||
Temperature fluctuations | ±0.5℃(at 200℃) | |||
Temperature distribution accuracy | ±2℃ (at 200℃) | |||
Heating time | The maximum temperature of 200°C arrives in about 60 minutes (to 200°C). | |||
Cleanliness | The temperature is kept at 100 level (0.5um particles) when the temperature is constant | |||
Constitute | Combinations | 6 slot combination | 4 slot combination | 2 slot combinations |
Liner material | SUS304# stainless steel plate | |||
Outer box material | The outer box material is mainly made of brushed stainless steel plate or SS41# carbon steel plate (smooth coating and baking paint after phosphoric acid film salt treatment), which can prevent fine dust | |||
Insulation | Aluminum silicate insulation cotton or ultra-fine glass wool | |||
heater | Stainless steel heating tubes | |||
Circulating motors | Centrifugal blades, high-temperature self-cooling motor 120W, 1 per tank | |||
Differential pressure gauge | Analog 0~500Pa | |||
Lead holes | Inner diameter 50mm (one on the right side) | |||
filter | Heat-resistant HEPA filter (dust collection efficiency: 0.3um particles 99.97% or more). | |||
N2 import interface | Outer diameter 8mm pagoda joint | |||
Controller | Temperature control | Japanese conductive temperature controller PID control mode | ||
timer | 1 minute ~ 99 hours 59 minutes and 100 ~ 999 hours 50 minutes (with timer waiting function) | |||
Run the function | Fixed-value run, program run | |||
Program mode | The program runs 4 sets of 10 segments (40 segments× 1, 20 segments× 2, 10 segments × 4). | |||
Additional features | Deviation correction function, key lock function, power failure compensation function | |||
sensor | PT100 platinum resistance (temperature regulator and overlift preventer). | |||
Safety devices | Leakage protection, phase sequence protection, temperature over-temperature protection, motor over-current protection, key lock function | |||
Specification | Single-slot inner size | 400×500×400 | 500×500×400 | 550×600×500 |
(D×W×H mm) | ||||
Outer dimensions | 1200×1880×1980 | 1250×1880×1980 | 1200×1000×1500 | |
(D×W×H mm) | ||||
The volume of a single slot | 80L | 100L | 150L | |
shelves | 2 floors | 2 floors | 2 floors | |
Heating power | 2 KW×6 | 2KW×4 | 3KW×2 | |
Power supply (50Hz) | Single-phase AC220V, 13KW | Three-phase AC220V, 9 KW | Three-phase AC380V, 7KW |